Radio Waves in Magnetic Resonance Imaging (MRI)

Radio waves in MRI

Author : Ms. Kalpana Parajuli
MSc Medical Imaging Graduate

It is true that we, who are associated with field of medical imaging, are mainly concerned with detrimental effects of ionizing electromagnetic radiation. It seems that our attention is biased and it is high time we realized we are surrounded by non-ionizing radiation and we should know how they interact with biological tissue. This article will mainly focus on radio waves, electromagnetic waves of frequency ranging from 9 kHz and 300 GHz that are used to perturb the longitudinal magnetization in order to produce MR signal. Those of us who belong to this dynamic and ever progressing field of medical imaging should know that the way by which radio wave interacts with the matter is quite different from the pathway that ionizing radiation such as x radiation takes. It is because these waves lack enough energy needed for compton or photoelectric interaction. For the ionization or the breakage of covalent bond to occur a single photon should interact with the electrons in the atomic orbital which is not possible with RF waves that releases its energy through the interaction of multiple photons. The effects of RF energy in human body can be divided into two categories.


a) Thermal also called dielectric heading:


 Heating effect of RF waves is best explained by the phenomenon polar molecules show under an electri field. Radio waves are nothing but the oscillating electric and magnetic fields. Under the sinusoidally changing electric field, the polar molecules, molecule with non-zero dipole moment,  begin to rotate, as a result of which collision between molecules occurs followed by energy transfer from these molecules to the adjacent molecules. The resultant agitation and energy transfer cause increase in temperature and production of heat respectively. Similarly changing magnetic field induces the electric field (Faradays law) in human body that deposits the energy in the form of heat as described above.


The behavior of electric and magnetic field of radio waves depends on the distance between the source of radiation and the object on which it is incident, as well as on its frequency. The patient in MRI is exposed to waves of frequency that  range from  8.5 to 340 MHz and the scanner is in the “near field” i.e “d” is less than one wavelength so heat production is mostly by the magnetic field of RF waves with little or minimal effect of electric field of RF wave. The reason behind this sort of dominance is the independent relationship between “E” and “B” in the near field.

 b) Non-Thermal :


These are specific effects that occur due to interaction of magnetic and electric field vector with human body other than heating. Very less is known about non-thermal effects because of lack of conclusive scientific evidence in human models and therefore not taken into account by regulatory bodies while providing guidelines on safety limit.
Specific Absorption Rate (SAR) is the name of dosimetric quantity used to describe the rate of heat deposition in unit mass of tissue, and is measured in W/kg.

What affects SAR in MR imaging?
The answer to this question is given by the figure below.



It is thus clear from above formula that SAR is the function of patient related factors as well as scan parameters. SAR increases with the square of magnetic field strength of the magnet system of scanner, because high power radio waves are needed to cause resonating effect on the spins of higher larmor frequency. It also increases with the square of the flip angle, whereas it varies proportionally with the duty cycle (ratio of average to peak RF power) and patient weight. Because conductivity and tissue density vary from person to person, the calculation of SAR is not easy and accurate. Animal experiments have shown that the permittivity and conductivity of tissue decreases with age thus young ones are more vulnerable to deleterious effects of radio waves. It also depends on the perfusion status and geometric configuration of exposed tissue and presence of metallic implants. Dependency of SAR on several factors is one reason behind the challenges in RF dosimetry.


SAR of 1W/kg is said to cause increase in temperature of insulated object (phantom) by 1⁰C in an hour. In human and animals, input of 4W/kg of SAR has shown to raise the temperature by 1⁰C. It is not practical to measure change in temperature (core/whole body or localized) thus SAR is used to quantify the RF exposure. International Electrotechnical Commission( IEC 60601-2-33) and Food And Drug Administration (FDA) has given limits for both temperature as well as values of SAR (shown below). The guidelines consist of limits for the whole body and local level exposure, because some organs are highly heat sensitive than others because of higher resistance, and are less affected by thermoregulatory system of body because of less perfusion, for example:  eye, gonads, and thus separate limit for local RF exposure was required.



It is now realized that separate guideline is necessary to account the change (increase) in SAR by metal implanted in the body of patients. However none of the current guidelines have addressed this issue. The limits are also given for occupational exposure. Those given by (IEC) are same as that for patients whereas that by Institute of Electrical and Electronics Engineers (IEEE) and International Commission on Non-Ionizing Radiation (ICNIRP) are one tenth of the maximum limit for patients.


Just like CTDI volume and Dose Length Product are displayed in the CT scanner, SAR value is also displayed on the monitor of the MRI system. There are many methods available for estimation of SAR. Two basic methods are caloriemetric method and pulse energy method. At the beginning of the scanning the machine runs calibration to find out the energy required to flip the spins by 90 and 180 degree. Power is then obtained by dividing the total energy of all pulse in one sequence with time of repetition (TR), and the result is ultimately divided by weight of the patient to get SAR. This is why the scanner requires us to input the value of weight of the patient.


Till now we believe that increase in patient weight increases the SAR and all of us have habit of looking at the value of SAR given by the machine itself to determine whether our protocols are safe. However results of a recent study were quite astounding because negative correlation was observed between patient weight and the SAR calculated by 3T scanner, whereas in 1.5 T scanners the relationship was maintained. This study has indeed raised a big question mark on the reliability of the values of SAR provided by the manufacturers of MRI scanner.


What about the consequences of RF induced heating in human?
Many incidents of second and higher degree burn in patients undergoing examination in 1.5 and 3.0T MR systems have been reported. While documentation of adverse consequence of RF associated heating of the metallic implants are available, other physiologic response for example- changes in heart rate, oxygen saturation, blood pressure, respiratory rate and cutaneous blood flow has not shown to cause effects that needs serious concern.


What are the possible ways to reduce SAR in MRI?
We are familiar with the tradeoffs between the radiation dose in CT and the image quality. A balance between them is essential part of protocol selection. Similarly in MRI as well, SAR can be minimized through wise selection of parameters, which however also affects other areas like imaging time, image quality etc. For example: reducing the flip angle affects the image contrast, reducing the no of slices increases imaging time. Other methods can be; reducing the echo train length of turbo spin echo or fast spin echoes, the use of quadrature rather than linear coils for
transmission. Parallel imaging technique, by using multiple receivers increases the amount of data received and thus reduces the imaging time by a certain factor (reduction factor) without the use of additional RF pulse and thus reduces SAR.


In conclusion safety issues regarding use of radio waves in MRI are gaining much more attention than ever mainly because of three reasons - 3T scanners, that needs high power RF amplifiers, about 35 kW, are widely being used; the popularity of new RF intensive sequences (HASTE, FIESTA, true FISP) have also heightened and MRI is no longer contraindicated for the patients with implanted materials. Thus it is necessary that manufacturer, the scientific research community and user should collaborate with each other to prevent the potential hazardous situation that can arise due to RF energy in MR environment.


MSc Medical Imaging Graduate Ms. Kalpana Parajuli
Maharajgunj Medical Campus (MMC),
Tribhuvan University Teaching Hospital (TUTH),
Kathmandu, Nepal



Friday, October 10, 2014

What is radiation quality?

What is radiation quality? 

A measure of the damage potential of a particular radiation type,this means alpha betas and gammas and positrons for example,has different abilities to cause damage to tissue,this damage potential is related to its size, charge and energy,we have a good example of this with amunitions, let's take a unit that you already know well,

energy gained by electron being accelerated through a field of 1 V,now to begin the radiation damage to tissue, let's first see the linear energy transfer,it is a measure of energy absorbed by the material, keV/length...somewhat related to the track or path.let's consider a beta particle moving through a particle such as lead brick,it should be intuitive and it interacts, it transfers it's energy to the material.high LET=more energy transfer and shorter distance travelled.Because it loses energy quickly.ability to produce more damage,of course we are concerned with biological damage,before we begin to understand the damage of radiation, we must discuss radiation dose measurements in radiation symmetry.

radioactivity rate of decay= number of disintegration per unit time (usually sec)
exposure=the electric charge in air due to the presence of radiation, the total sum of kinetic energy of charged particles liberated by uncharged radiation.

absorbed dose=similar to exposure-consider tissue type
energy deposited in tissue.value of absorbed dose from same radiation will change with tissue type.

1Roentgen = 1Rad
100rad=1Gy
in human tissue, about 87% of 1R is absorbed 0.87R=1rad
dose equivalent=corrects absorbed dose for damage potential of radiation type
must provide a weight factor of radiation types.
QF or RBE,gives numerical value to the damage potential.
QF=dose from standard radiation to produce given biological effect/dose from test radiation to produce same biological effect
standard radiation -200 kVp x-rays
dose equivalent:
rem \ sievert
100 rem=1Sv,
Dose equivalent =absorbed dose x QF or RBE
ICRP defined values for QF/RBE
gamma x rays 1
electrons (beta/positron) 1
neutrons 10
alpha 20
QF INCREASES as LET increases
150mrad x 1 gamma - 150 rem
75 mrad x 20 alpha = 1500 rem
total dose equivalent = 150 + 1500 rem
for QF =1
general rule 1 R =1 rad = 1 rem
reality:
1 rad = 0.96 rem or 0.87 R = 1 rad
there is no universally accepted conversion from rad to rem.
exposure: air
absorbed dose: considers material
dose equivalent: considers radiation type
effective dose equivalent:
measures the biological risk on the whole body by considering different tissue types response differently to radiation ; also accounts for radiation to only part of the body vs the whole body.

to account for different tissue response, weight factors are used (yes, again), weight factors determined by ICRP.
2007 revision of ICRP recommendations:
14 designated tissues for organs + 1 for everything else.
to calculate: summation of- Dose Equivalent x Weight Factor
For all designated tissues/organs irradiated
H = summation Wt x Ht
How do we account for long term dose to an organ? inhaled, injected or swallowed
committed dose equivalent: Ht,50, the dose equivalent to an organ or tissue received during the 50 year period following intake of the radioactive material, we use 50 because there isnot so much damage after 50,as the receipent would likely to be perished by then.committed dose equivalent accounts for physical decay and body excretion.

biological half life: SAME principle as radioactive or physical half life,and it is how we account for body excretion of radioactive material,it is important to know that it is an approximation differs from person to person,sickness are different and medications are different,

Biological Half life:depends on the renal system and the bowels, the overall disappearance of radioactivity from the body depends on effective half life, consider both physical and biological decays,
1/effective halflife=1/biological halflife + 1/physical halflife
we have seen many ways to calculate dose now considering many variables.
it becomes very complex trying to calculate a specific dose to a specific organ.
source organ is lung, target organ is liver.after.....then same organ liver can be source and target.
to calculate dose to the liver precisely we need to answer few questions//
exact location in lungs, time spent in lungs, exact location lin liver, time spent in liver.
doses from each adjacent organ of liver must be considered, which becomes even more complicated.
another factor we must consider in calculating dose to tissue is the percentage of radioactive emissions that deposit energy in the tissue vs those that escape.this value is called the absorbed fraction.
depends on track, density and thickness, radiation type, energy and half life.
absorbed fraction will be expressed as a number b/w 0-1, where 1.0 equals 100% absorption.
with all of these factors, how can we ever calculate an accurate dose to an organ?
MIRD - Medical Internal Radiation Dose Committee:which is the permanent committe of nuclear medicine provided a widely accepted method to calculate internal dose,
the first method is to calculate cumulated activity;plot activity vs time, area under the curve is the cumulated activity;units = activity x time uCi x hr
it is important to understand that the cumulated activity is the first step in determining the dose to an organ that a patient receives from a nuclear medicine procedure.
S factor: consider most of the physical aspects of location of radioactive material,
total energy associated with the radiation typexabsorbed fraction in target of source/mass of the target organ.
complications of calculating S factor: absorbed fraction varies greatly with gamma rays, also all transitions, disintegrations and emissions must be included not just the primary decay, for 99mTc, there are altogether 14 radiations to include.this include conversion electrons, fluorescent x-rays and auger electrons,

lucky for us, MIRD published all the values of S-factor, so we don't need to calculate all these values.
DOSE equation of MiRD=cumulated activity x S-factor
Cumulated Activity = Biological Parameters
S-factor = Physical parameters
 review:
LET and its affect on dose
all measurements of radioactive energy deposit
unit conversion b/w SI and traditional
how to incorporate QF in dodse calculations
how to calculate effective half life
how to calculate organ dose using MIRD scheme

Reference: